The gamma irradiation device
GSR D1 is the optimal solution
for gamma irradiation in the radiobiological
research. The maximum
irradiation capacity of 64
l in connection with a low surface
dose rate offers efficiency
and safety in your laboratory.
Volume
The irradiation chamber dimensions of 400 x
400 x 400 mm with three adjustable irradiation
levels. The GSR D1 has secure access for cables
and instrumentation.
Dose Rate
The activity of Cs-137 sources can be offered
according to the users specifications. A dose
difference of < 10 % can be realised on irradiation
level two.
Operation
The operation of the GSR D1 is user friendly because
of its high quality and reliable components
as well as its comprehensive safety system.
The current status is indicated on the display.
Radiation safety
GSR D1 gamma irradiation devices are selfcontained
devices with a surface dose rate of
< 5 μSv/h with its maximum activity of 200 TBq
Cs-137.
Safety Feature
- Operation with key switch and safety code
- Irradiation cycle will continue for up to 300 min, in case of power failure
- Independent locking of loading hatch and irradiation chamber
- Continuous monitoring and control of the operation during irradaition
- Self-check of the irradiation unit before and during the irradiation procedure
Technical Spezification
| Irradiation Device |
|
| Weight |
4300 kg |
| Unit Dimensions |
Height 1520 mm |
| |
Width 1210 mm |
| |
Depth 690 mm |
| Recommended room size |
min. 2 x 2 m |
| Electrical supply |
115 V / 230 V |
| Room temperature |
+ 15 until + 35° |
| Battery Back-up |
Yes |
| Certificate |
CE |
| Irradiation Chamber |
| Measurements |
400 x 400 x 400 mm |
| Volume |
64 litre |
| Cable acces |
2 |
| Air input access |
3 |
| Source |
| Source Activity |
Cs-137 up to 200 TBq |
| Surface dose rate |
< 5 μSv/hr |
| Dose uniformity |
< 10 % over a 300 mm square |
| Accessories |
| Collimators on request |
|
| Attenuators on request |
|
| Rodent holder |
|